You are here: Home » Blog » Guide » Silicone Shore A Hardness Explained: How to Select the Right Durometer for Your Medical Application

Silicone Shore A Hardness Explained: How to Select the Right Durometer for Your Medical Application

Views: 0     Author: Kevin Fang     Publish Time: 2026-06-09      Origin: Chensheng Medical

Table of Contents

If you have ever reviewed a silicone tubing datasheet or submitted a custom OEM inquiry, you have encountered the term Shore A hardness — typically expressed as a number between 20 and 80, followed by the letter "A." It appears on every technical data sheet, every material certificate, and nearly every product drawing for silicone components.

Yet it is one of the most commonly misunderstood and incorrectly specified parameters in medical silicone procurement. Engineers routinely specify Shore A values that are either too soft for the application (causing kinking, collapse, or dimensional instability) or unnecessarily firm (reducing patient comfort, increasing pump wear, or making assembly difficult).

This guide gives you a complete, practical understanding of Shore A hardness — what it measures, how it is tested, how it interacts with other design parameters, and how to select the right value for every major medical silicone application.

Silicone Shore A Hardness Explained: How to Select the Right Durometer for Your Medical Application

What Is Shore A Hardness? The Physical Meaning

Shore A hardness is a standardized measure of a material's resistance to permanent indentation by a defined probe under a defined load. It is one of several Shore hardness scales developed by Albert Ferdinand Shore in the 1920s, each designed for different material stiffness ranges.

For silicone rubber — and elastomers in general — the Shore A scale is the standard measurement tool. The test is defined by ASTM D2240 (Standard Test Method for Rubber Property — Durometer Hardness) and its international equivalent ISO 48-4.

How the Test Works

A Shore A durometer consists of a truncated cone-shaped indenter with a flat tip (0.79mm diameter), loaded by a calibrated spring. The indenter is pressed perpendicularly into the flat surface of the test specimen under a total force of 822 grams (8.06 N). The depth of penetration after a defined contact time (typically 15 seconds for time-dependent readings) is converted to a hardness value on a scale of 0 to 100:

  • 0 = the indenter penetrates to its full depth (infinitely soft material)

  • 100 = zero penetration (infinitely hard material — equivalent to steel)

In practice, medical silicone tubing and molded components span the range from approximately Shore A 10 (very soft, gel-like) to Shore A 80 (firm, stiff rubber).

Test Specimen Requirements

For a valid Shore A measurement per ASTM D2240:

  • Minimum specimen thickness: 6.4mm (0.25 inches)

  • Flat, parallel surfaces required

  • Temperature: 23°C ± 2°C (results are temperature-dependent)

  • Minimum distance from edge: 12mm

Critical note for tubing: Shore A hardness is measured on flat slab specimens of the compound — not on the tube itself. A thin-wall tube cannot be directly measured with a durometer. The value on a tubing datasheet refers to the compound used to make the tube, tested on a flat slab. This is why wall thickness and Shore A hardness must always be considered together when evaluating tubing flexibility.

Shore A vs. Shore D vs. Shore OO: Which Scale Applies?

The Shore system includes multiple scales for different material types. Understanding which scale applies to your material prevents specification errors:

Scale

Indenter Type

Load

Applicable Materials

Typical Range

Shore OO

Sphere (1.19mm radius)

113g

Ultra-soft gels, foams, very soft elastomers

Silicone gels: 5–30 OO

Shore A

Truncated cone (flat tip)

822g

Soft to medium rubbers, elastomers, silicone

Medical silicone: 10–80 A

Shore D

Sharp cone (30° tip)

4,536g

Hard rubbers, rigid plastics, hard elastomers

Hard silicone: 80+ A / 20–80 D

For medical silicone tubing and molded components, Shore A is always the correct scale. Shore D is used for hard thermoplastics (ABS, PC, nylon) used in device housings and rigid components — not for silicone.

The scales overlap at the extremes: a material measuring Shore A 90–100 can also be measured on the Shore D scale. A material measuring Shore A 10–20 may be better characterized on the Shore OO scale for precision. For the vast majority of medical silicone applications (Shore A 20–80), the Shore A scale provides clear, unambiguous measurements.

The Shore A Hardness Spectrum for Medical Silicone

Medical silicone is available across a wide hardness range. Here is what each range feels like and where it is used:

Shore A 10–20: Ultra-Soft / Gel-Like

Physical feel: Barely perceptible resistance to touch; deforms under the weight of a finger; recovers slowly.

Material behavior:

  • Extremely high elongation (600–800%)

  • Very low tear resistance — handle with care

  • Excellent conformability to irregular surfaces

  • Prone to dimensional instability under any load

Medical applications:

  • Neonatal skin-contact components (pressure relief)

  • Soft tissue simulation and training models

  • Implant-adjacent cushioning components

  • Specialty wound contact interfaces

Not suitable for: Any application requiring dimensional stability, fluid containment under pressure, or pump contact.

Shore A 20–35: Very Soft / Highly Flexible

Physical feel: Noticeably soft; easily deformed with light finger pressure; quick elastic recovery.

Material behavior:

  • High elongation (500–700%)

  • Moderate tear resistance

  • Excellent flexibility — resists kinking at very small bend radii

  • Low compression set resistance — not suitable for dynamic compression

Medical applications:

  • Neonatal and pediatric feeding tubes (patient comfort priority)

  • Soft drainage tubes for sensitive anatomical locations

  • Gentle patient-contact tubing (nasal, oral)

  • Balloon components requiring high compliance

  • Specialty catheter tips for atraumatic insertion

Chensheng Medical note: Our neonatal feeding tube range uses Shore A 25–30 platinum-cured compound to achieve the combination of softness and dimensional precision required for NICU applications. Learn more about our medical silicone tubing product range.

Shore A 35–50: Soft / Standard Medical Flexibility

Physical feel: Soft and pliable; deforms easily under moderate finger pressure; springs back quickly.

Material behavior:

  • Good elongation (400–600%)

  • Good tear resistance

  • Excellent flexibility for general fluid transfer

  • Adequate compression set for low-frequency dynamic applications

Medical applications:

  • General-purpose IV fluid transfer tubing

  • Standard urological drainage tubing

  • Enteral feeding tubes (adult)

  • Nasogastric tube bodies

  • Soft catheter shafts

  • General medical fluid handling tubing

This is the most widely used hardness range for standard medical tubing. It represents the best balance of flexibility, dimensional stability, and ease of assembly for the majority of fluid transfer applications.

Shore A 50–65: Medium / Resilient — The Pump Tubing Zone

Physical feel: Noticeably firmer; requires deliberate force to deform; snaps back immediately upon release.

Material behavior:

  • Good elongation (350–500%)

  • High tear resistance

  • Excellent compression set resistance (≤8% at 22h/70°C)

  • Optimal fatigue resistance under cyclic compression

  • Good dimensional stability under pressure

Medical applications:

  • Peristaltic pump tubing (the primary application for this range)

  • Respiratory circuit tubing (adult ventilator circuits)

  • Anesthesia breathing circuit components

  • High-flow fluid transfer tubing

  • Pump segments for infusion pumps

  • Tubing requiring resistance to external compression

Why Shore A 50–65 is critical for peristaltic pump tubing: Peristaltic pumps work by repeatedly compressing and releasing the tubing. The tubing must recover its full circular cross-section between roller contacts to maintain accurate flow rates. Too soft (below Shore A 45) and the tube recovers slowly, causing flow rate drift and eventual collapse. Too firm (above Shore A 70) and the pump motor must work harder, increasing wear on both the tube and the pump head. Shore A 55–62 is the optimal range for most peristaltic pump applications.

For a complete technical guide to peristaltic pump tubing selection — including compression set testing, fatigue life, and pump head compatibility — see: Peristaltic Pump Tubing Selection: Material Properties and Performance Factors

Shore A 65–80: Firm / Structural

Physical feel: Distinctly firm; requires significant force to deform; minimal visible deflection under hand pressure.

Material behavior:

  • Moderate elongation (200–400%)

  • Very high tear resistance

  • Excellent pressure rating

  • High dimensional stability

  • Reduced flexibility — minimum bend radius increases significantly

Medical applications:

  • High-pressure fluid transfer tubing

  • Structural silicone components in device assemblies

  • Silicone gaskets and seals requiring high compression resistance

  • Connector bodies and rigid interface components

  • Tubing requiring resistance to internal pressure (>3 bar)

  • Silicone O-rings for static sealing applications

Caution: At Shore A 70+, silicone tubing becomes noticeably stiff. For patient-contact or flexible fluid transfer applications, this range is rarely appropriate. Verify that the stiffness is genuinely required before specifying above Shore A 65.

Silicone Shore A Hardness Explained: How to Select the Right Durometer for Your Medical Application

The Critical Interaction: Shore A Hardness × Wall Thickness

This is the most important concept in silicone tubing specification — and the most frequently overlooked.

Shore A hardness measures the material. Wall thickness determines how that material behaves as a tube.

Two tubes can have identical Shore A hardness but completely different flexibility if their wall thicknesses differ. Conversely, two tubes can feel equally flexible despite having different Shore A values if their wall thicknesses are adjusted accordingly.

The Stiffness Relationship

The bending stiffness of a tube scales approximately with the cube of wall thickness for thin-walled tubes:

Bending Stiffness∝E×t3Bending Stiffness∝E×t3

Where EE is the elastic modulus (directly related to Shore A hardness) and tt is wall thickness.

This means: doubling the wall thickness increases bending stiffness by approximately 8× — far more than the effect of changing Shore A hardness within the normal medical range.

Practical Implications

Scenario

Shore A

Wall Thickness

Result

Standard IV tubing

45

1.0mm

Flexible, easy to handle

Same compound, thicker wall

45

2.5mm

Noticeably stiffer — may be too stiff for IV use

Firmer compound, thin wall

60

0.8mm

Similar flexibility to the first example

Pump tubing

60

1.6mm

Firm enough for compression recovery, flexible enough for routing

High-pressure tubing

70

3.0mm

Very stiff — intentional for pressure containment

Design rule: When you need a tube that is both flexible (for routing and patient comfort) AND dimensionally stable (for accurate flow rates or pressure containment), the solution is often a thinner wall in a slightly firmer compound — not a softer compound with a thicker wall. Discuss this trade-off with your supplier's applications engineering team before finalizing your specification.

Shore A Hardness and Key Mechanical Properties: What Changes

When you change Shore A hardness, you change multiple mechanical properties simultaneously. Here is how the key properties shift across the medical silicone hardness range:

Property

Shore A 20–30

Shore A 40–50

Shore A 55–65

Shore A 70–80

Tensile strength (MPa)

4–6

7–9

8–11

9–12

Elongation at break (%)

600–800

400–600

350–500

200–400

Tear resistance (kN/m)

10–18

20–28

25–35

30–40

Compression set (%, 22h/70°C)

15–25

10–18

6–12

5–10

Flexibility

Excellent

Very good

Good

Moderate

Kink resistance

Low

Moderate

Good

Very good

Pressure rating

Very low

Low–moderate

Moderate

High

Pump suitability

Poor

Marginal

Excellent

Poor (too stiff)

Patient comfort

Excellent

Very good

Good

Poor

Values are typical ranges for platinum-cured medical silicone. Actual values depend on specific compound formulation.

Why Compression Set Matters for Pump Tubing

Compression set is the percentage of deformation that remains after a compression load is removed. A compression set of 10% means the tube does not fully recover 10% of its original diameter after being compressed.

For peristaltic pump tubing, compression set directly determines:

  1. Flow rate accuracy over time — as compression set accumulates, the effective ID decreases, reducing flow rate

  2. Tube service life — high compression set leads to permanent deformation and eventual tube collapse

  3. Pump calibration stability — pumps calibrated at installation drift as compression set increases

This is why platinum-cured silicone is non-negotiable for pump tubing — its compression set values (6–10%) are significantly better than peroxide-cured silicone (15–25%). For a detailed comparison of curing chemistry and its effect on mechanical properties, see: Platinum-Cured vs. Peroxide-Cured Silicone: Which Is Better for Your Application?

Application-to-Hardness Quick Selection Guide

Use this table as your starting point for Shore A specification. Always validate with functional testing in your specific application.

Application

Recommended Shore A

Wall Thickness Guidance

Key Rationale

Neonatal feeding tube

20–30

0.5–0.8mm

Patient comfort, atraumatic insertion

Adult nasogastric (NG) tube

30–40

0.8–1.2mm

Comfort + sufficient stiffness for placement

PEG / gastrostomy tube

40–55

1.0–1.5mm

Durability for long-term indwelling use

IV fluid transfer tubing

40–55

1.0–1.5mm

Flexibility, transparency, easy assembly

Urological drainage tubing

35–50

1.0–1.5mm

Patient comfort, kink resistance

Wound drainage tubing

40–55

1.2–2.0mm

Kink resistance, tissue compatibility

Peristaltic pump segment

55–65

1.2–2.0mm

Compression set, fatigue life, flow accuracy

Infusion pump segment

55–62

1.4–1.8mm

Compression recovery, calibration stability

Adult ventilator circuit

55–65

1.5–2.5mm

Kink resistance, low compliance, durability

Pediatric ventilator circuit

55–65

1.2–2.0mm

Kink resistance, lower dead space

Neonatal ventilator circuit

60–70

1.0–1.5mm

Dimensional stability, minimal dead space

Anesthesia breathing circuit

55–65

1.5–2.5mm

Low compliance, kink resistance, durability

High-pressure fluid transfer

65–75

2.5–5.0mm

Pressure containment

Silicone O-ring / gasket (static)

60–70

N/A (molded)

Sealing force, compression set

Silicone O-ring / gasket (dynamic)

50–60

N/A (molded)

Wear resistance, sealing under motion

Balloon catheter balloon

20–35

0.3–0.6mm

High compliance, low inflation pressure

Foley catheter shaft

40–55

1.0–1.5mm

Comfort, kink resistance, durability

Surgical drain (Penrose type)

30–45

0.5–1.0mm

Tissue compatibility, flexibility

Pharmaceutical fluid transfer

45–60

1.0–2.0mm

Low extractables, dimensional stability

Micro-bore analytical tubing

50–65

0.3–0.8mm

Dimensional precision, chemical inertness

For respiratory circuit applications, see our detailed compliance guide: Medical-Grade Silicone Tubing for Respiratory Circuits: Compliance Requirements

How Shore A Hardness Is Specified in an OEM Order

When placing a custom OEM order for medical silicone tubing or molded components, Shore A hardness should be specified as follows:

Option 1: Nominal Value with Tolerance (Preferred)

Specify a target value with an acceptable tolerance band:

  • Example: Shore A 60 ± 5

  • This allows the manufacturer to use a qualified compound within the acceptable range

  • Tolerance of ±5 Shore A units is standard; ±3 is achievable for precision applications

Option 2: Range Specification

Specify an acceptable range:

  • Example: Shore A 55–65

  • Equivalent to Option 1 but expressed as a range

  • Useful when the application has a clear performance window

Option 3: Performance-Based Specification (Advanced)

For critical applications, specify the performance requirement rather than the hardness:

  • Example: "Compression set ≤ 8% per ASTM D395 Method B, 22h at 70°C"

  • This allows the manufacturer to optimize the compound for the performance outcome

  • Requires incoming inspection testing to verify compliance

Best practice: For peristaltic pump tubing, always specify both Shore A hardness AND compression set — hardness alone does not guarantee pump performance. Two compounds at Shore A 60 can have very different compression set values depending on their formulation and curing system.

What to Include in Your Technical Drawing

A complete silicone tubing drawing should specify:

  1. Inner diameter (ID) with tolerance

  2. Outer diameter (OD) with tolerance

  3. Wall thickness (WT) with tolerance — note: only two of three are independent

  4. Shore A hardness with tolerance

  5. Color (transparent / specified color)

  6. Curing system (platinum-cured — always for medical)

  7. Applicable certifications (USP Class VI, ISO 10993, FDA 21 CFR 177.2600)

  8. Special requirements (radiopaque stripe, cleanroom packaging, sterilization)

For guidance on the complete custom OEM development process — from drawing submission through tooling, sampling, and production approval — visit our medical silicone products page or contact our applications engineering team directly.

Measuring Shore A Hardness: What to Verify on Incoming Inspection

When receiving medical silicone tubing from any supplier, Shore A hardness should be verified as part of your incoming inspection protocol.

Practical Measurement on Tubing

Because tubing cannot be directly measured with a standard durometer (insufficient thickness, curved surface), there are two approaches:

Approach 1: Request compound slab test specimensAsk your supplier to provide flat slab specimens (minimum 6.4mm thick) of the same compound batch used for your tubing. Measure per ASTM D2240 at 23°C ± 2°C.

Approach 2: Verify against lot-specific CoARequest a lot-specific Certificate of Analysis (CoA) that includes the actual Shore A measurement for the compound lot used in your production run. Verify the reported value is within your specified tolerance.

Approach 3: Micro-durometer for thin sectionsFor thin-wall tubing where slab specimens are not available, a micro-durometer (Shore M scale) can be used on the tube wall directly. Note that micro-durometer values are not directly equivalent to standard Shore A values — establish a correlation factor with your supplier.

Red Flags in Shore A Verification

  • CoA shows only nominal value, not actual measured value: The supplier may be copying the compound specification rather than testing each lot. Request actual test data.

  • Shore A value outside specification by more than 3 units: Indicates compound formulation inconsistency or wrong compound used. Initiate supplier CAPA.

  • Shore A correct but compression set high: Indicates peroxide-cured compound may have been substituted. Request curing system confirmation.

For guidance on evaluating supplier quality systems and documentation, see: How to Choose a Reliable Medical Silicone Manufacturer in China

Common Specification Mistakes — and How to Avoid Them

Mistake 1: Specifying Shore A Without Wall Thickness

Problem: Shore A alone does not define the flexibility of a tube. A Shore A 60 tube with 0.8mm wall is more flexible than a Shore A 40 tube with 3.0mm wall.Solution: Always specify Shore A and wall thickness together. If flexibility is the key requirement, specify the minimum bend radius instead.

Mistake 2: Copying a Competitor's Shore A Value Without Understanding Why

Problem: You see a competitor's product listed as "Shore A 55" and specify the same value without understanding whether it is appropriate for your application.Solution: Use the application-to-hardness table above to verify the specification makes sense for your use case. If uncertain, request samples at Shore A 50, 55, and 60 and test functionally.

Mistake 3: Specifying Shore A 40 for Pump Tubing

Problem: Shore A 40 feels soft and flexible — which seems desirable. But it has poor compression set recovery, leading to flow rate drift and premature tube failure in peristaltic pump applications.Solution: For peristaltic pump tubing, specify Shore A 55–65 regardless of how firm it feels. The compression set performance, not the feel, is what matters.

Mistake 4: Specifying Shore A 70+ for Patient-Contact Tubing

Problem: High Shore A was specified for "durability" or "pressure resistance" without considering that the resulting stiffness is uncomfortable for patients and may cause tissue trauma.Solution: For patient-contact applications, Shore A 35–55 is almost always sufficient. Use wall thickness and reinforcement (braid, spiral) to achieve pressure resistance rather than increasing hardness.

Mistake 5: Not Accounting for Temperature Effects

Problem: Shore A hardness decreases as temperature increases. A tube specified at Shore A 60 at 23°C may measure Shore A 50–55 at body temperature (37°C) and Shore A 40–45 at autoclave temperature (121°C).Solution: For applications where operating temperature differs significantly from 23°C, request hardness measurements at the actual operating temperature, or apply a correction factor.

Frequently Asked Questions (FAQ)

Q1: What Shore A hardness should I specify for standard medical silicone tubing?

A: For general-purpose medical fluid transfer tubing (IV lines, drainage, general fluid handling), Shore A 40–55 is the standard starting point. This range provides the best combination of flexibility, dimensional stability, and ease of assembly for the majority of medical applications. If you have a specific application — pump tubing, respiratory circuits, neonatal use — refer to the application-to-hardness table above or contact our applications engineering team for a specific recommendation.

Q2: Is there a difference between Shore A hardness and durometer hardness?

A: "Durometer hardness" is the general term for hardness measured with a durometer instrument. "Shore A hardness" specifies both the instrument type (Shore durometer) and the scale (A scale). In practice, the terms are used interchangeably for elastomers. When someone says "durometer 60" without specifying the scale, they almost always mean Shore A 60 for silicone rubber applications.

Q3: Can I get silicone tubing in any Shore A value, or are there standard grades?

A: Silicone compounds are available in a wide range of Shore A values, but manufacturers typically stock standard grades (commonly Shore A 30, 40, 50, 60, 70). Custom hardness values between these grades are achievable but may require a minimum order quantity to justify compound preparation. At Chensheng Medical, we maintain stock compounds at Shore A 30, 40, 50, 55, 60, and 70 for our medical tubing range. Custom hardness values are available for OEM orders with sufficient volume.

Q4: How does Shore A hardness affect the pressure rating of silicone tubing?

A: Shore A hardness has a secondary effect on pressure rating — wall thickness is the primary determinant. However, firmer compounds (Shore A 65–80) do provide higher burst pressure for a given wall thickness because of their higher elastic modulus. For high-pressure applications (>3 bar), we recommend combining adequate wall thickness with Shore A 65–75. For a given pressure requirement, our engineering team can calculate the minimum wall thickness and recommend the appropriate hardness.

Q5: Does Shore A hardness change after sterilization?

A: For platinum-cured silicone, Shore A hardness is highly stable across all standard sterilization methods. Autoclave cycling (up to 200 cycles at 134°C) typically produces less than 3 Shore A units of change. Gamma irradiation at standard doses (25–50 kGy) may produce a slight hardening effect (2–5 Shore A units) due to additional crosslinking. Peroxide-cured silicone shows greater post-sterilization property changes — another reason platinum-cured compounds are preferred for medical applications.

Q6: We are qualifying a new silicone tubing supplier. How do we verify Shore A hardness compliance?

A: Request lot-specific CoA data showing actual measured Shore A values (not just nominal compound specification). For critical applications, request compound slab specimens from the same production lot and verify independently per ASTM D2240 at 23°C ± 2°C. Include Shore A verification in your incoming inspection protocol with defined acceptance criteria (e.g., nominal ± 5 Shore A units). For guidance on the full supplier qualification process, see our article on how to choose a reliable medical silicone manufacturer in China.

Q7: What is the Shore A hardness of Chensheng Medical's standard medical tubing grades?

A: Our standard medical silicone tubing is available in Shore A 30, 40, 50, 55, 60, and 70. Our most commonly specified grades for medical device applications are Shore A 50 (general fluid transfer, catheters, drainage) and Shore A 60 (peristaltic pump tubing, respiratory circuits). All grades use platinum-cured compound manufactured in our ISO Class 7 cleanroom under ISO 13485 quality management. Contact our team for samples of any grade for your qualification testing.

Silicone Shore A Hardness Explained: How to Select the Right Durometer for Your Medical Application

Summary: Shore A Hardness Selection in Five Steps

  1. Identify your primary application — pump tubing, patient-contact, structural, fluid transfer, or respiratory

  2. Determine the dominant performance requirement — flexibility, compression set recovery, pressure rating, or patient comfort

  3. Use the application-to-hardness table to identify the recommended Shore A range

  4. Consider wall thickness together with Shore A — they interact to determine actual tube flexibility and stiffness

  5. Request samples at two or three hardness values within the recommended range and validate functionally in your application before finalizing specification

When in doubt, our applications engineering team at Chensheng Medical is available to review your drawing, discuss your application requirements, and recommend the optimal Shore A specification — at no charge, as part of our standard pre-sales technical support.

Request Samples or a Custom Quotation

Chensheng Medical manufactures platinum-cured medical silicone tubing across the full Shore A range (20–80) in our ISO 13485-certified, ISO Class 7 cleanroom facility. All products are supported by full biocompatibility documentation (USP Class VI, ISO 10993, FDA 21 CFR 177.2600) and lot-specific Certificates of Analysis.

Contact Our Applications Engineering TeamBrowse Our Medical Silicone Tubing RangeRequest Free Samples for Qualification Testing

Related Articles:

Chensheng – China’s Leading Silicone Product Manufacturer

Choose Chensheng, and gain a trusted partner with over 20 years of OEM/ODM expertise. We deeply understand your needs and deliver professional, reliable, and tailored silicone solutions.

Quick Links

Get in Touch

Contact us

Copyright © 2025 JINAN CHENSHENG MEDICAL TECHNOLOGY CO., LTD.   鲁ICP备2021012053号-1      互联网药品信息服务资格证书 (鲁)-非经营性-2021-0178    中文站